Specification

  Specification

Eco-X

Eco-X Specifications

Eco-X Specifications

Scroll right on tables to view all table data.

Model Panorama Cephalo CT FOV (DxH) 16 x 9 cm Free FOV Model Scan
Eco-X x x x x x
Eco-X S x x (Scan type) x x x x
Mode Normal Low Dose
Scan Time (Adult) CBCT 24 s 8 s or 12 s
PANO Teeth Full Arch 14 s 7 s
Bitewings 8 s 6 s
Bitewing Right (or left) 4 s 3.2 s
TMJ Open / Close (Capture Both) 6 s 3.2 s
Open / Close (Select One) 3.2 s 1.7 s
CEPH Lateral Large Area 8 s 4 s
Lateral Small Area 6 s 3.2 s
Frontal(PA)/Carpus/ Waters/SMV 8 s 4 s
Model Scan 24s
Mode Normal Low Dose
Scan Time (Child) CBCT 24 s 8 s or 12 s
PANO Teeth Full Arch 12.7 s 6.5 s
Bitewings 6 s 4 s
Bitewing Right (or left) 3.2 s 2.2 s
TMJ Open / Close (Capture Both) 4 s 2.2 s
Open / Close (Select One) 2.2 s 1.2 s
CEPH Lateral Large Area 6 s 3.4 s
Lateral Small Area 3.4 s 2.6 s
Frontal(PA)/Carpus/ Waters/SMV 7 s 3.7 s
Model Scan 24s
Field of View (CT)
(cm × cm)
(Diam. × Height)

16 × 9 or 12 x 9(Adult)
10 × 8 (Child)
Free FOV (Min. 3 x 3)

Voxel Size 0.10, 0.15, 0.20, 0.25, 0.30
Focal Spot (mm) 0.5
Reconstruction Time min. 10 s (standard recon, MAR off)
Gray Scale CT & Panorama 16 bits
Scan- type Cephalo 14 bits
Tube Voltage 60 kV to 90 kV
Tube Current 4 mA to 10 mA
Patient Position Standing (Wheelchair accessible)
Detector Type CT & Panorama TFT: a-Si
Scan Type Cephalo CMOS
System Weight (Tolerance: 10 %) Eco-X: 178 kg/ 392 lbs
Eco-X S: 211 kg/ 465 lbs
Dimensions mm × mm × mm
(in × in × in) (W × D × H) (Tolerance: 5 %)
Eco-X: 980 × 1225 × 2314 (38.6 ” × 48.3 ” × 91.1 ”)
Eco-X S:1802 × 1225 × 2314 (71 ” × 48.3 ” × 91.1 ”)
Control Device Alignment : Touch Panel
Exposure & Imaging Acquisition: Workstation
Patient Alignment Column: Electric motion
Temple Support: Motion
Patient Position Standing (Wheelchair accessible)
Install Standing

Product Dimensions

Eco-X